Water Survey Tech

Product Detail

Multi-parameter probe MPS-D8/Qualilog8

  • Robust, pluggable multi-parameter sensor from 1.4404 stainless steel to accommodate up to eight sensors (water level, temperature, conductivity, pH, redox potential, oxygen, turbidity, etc.)
  • Additional electric contact gauge function combined with KLL-Q-2 mobile field laboratory
  • Slim design (Ø 48mm, 493 mm length), can be used with 2″ and larger pipe diameters
  • Output: RS485 (SHWP), Option: MODBUS, SDI-12, 0/4-20 mA
  • Optional: Integrated, powerful SEBA data logger Qualilog8 for stationary use
Categories ,

Share This Items :

Individually configurable:
Depending on the measuring task at hand, the MPS-D8 can be individually equipped with a maximum of eight water quality sensors and can detect up to 13 parameters.

Simple handling:
Special bayonet locks grant the users a simple and quick replacement of worn sensors. The calibration and configuration of the MPS-D8 is carried out using a special USB-RS485 interface cable together with ConfigPC operating software (Windows). This will automatically detect if a probe is spent and must be replaced.

Robust:
The housing of the MPS-D8 is made of high-quality 1.4404 stainless steel. The probe can thus be used even under adverse ambient conditions (e.g. landfill or other contaminated sites).

Variable:
The proven and robust SEBA plug system of the MPS-D8 allows both the connection to mobile handsets (SEBA KLL-Q-2 Type electric water level gauge (KLL)Checker-2) and stationary data collectors with data transmission (UniLogComLogCom/FlashCom) or without data transmission (Qualilog8).

  • Water quality monitoring
  • In surface waters: rivers, canals, springs, lakes
  • Groundwater: extraction wells, field measurement sites, landfill and construction sites
  • Other applications: discharge monitoring
Data collector:

External:

Internal: Qualilog-8

Electric water level gauge (KL):

KLL-Q-2

Handset:

Checker-2

Calibration:

Dual converter (USB/RS232 – RS485)

Operating software:

SEBAConfig (Windows)

Flow body
Parameters:

Water level [cm], [m]

Sensor:

Relative pressure sensor

Measurement principle:

Capacitive with ceramic measurement cell

Measuring range:

0-2 / 1 /4 /5 /20 /40 / 100/ 200m water column or according to specifications

Measuring accuracy:

± 0.1% FS

Long-term stability:

± 0.1% / year

 

Parameters:

Water temperature [°C]

Sensor:

NTC30 with polynomial linearisation

Measurement principle:

NTC resistance

Measuring range:

-5 – +50°C

Measuring accuracy:

0.1°C

Dissolution:

0.01°C

Pressure Range:

0-50 bar

 

Parameters:

Conductivity [µS/cm], [mS/cm]

Derived parameters: solutes [g/l], salinity, density [g/l]

Sensor:

4-electrode measuring cell

Measuring range:

0-200mS/cm (total)

Automatic measuring range switchover: 0-200µS/cm; 0,2-2mS/cm; 2,2mS/cm; 20-200mS/cm

Measuring accuracy:

± 1µS/cm for measuring range of 0-200µS/cm +/- 0,5% for measuring range of 0,2-200mS

Dissolution:

0.0001mS/cm

Temperature:

-5°C-50°C

Pressure Range:

0-50 bar

 

Parameters:

pH value [pH]

Sensor:

Glass electrode

Measuring range:

0-14pH

Measuring accuracy:

± 0.1pH

Dissolution:

0.01 pH

Temperature:

0-50°C

Pressure Range:

0-20 bar

 

Parameters:

Redox [mV]

Sensor:

Glass electrode

Measuring range:

-1200mV-1200mV

Measuring accuracy:

± 10mV

Dissolution:

0.1mV

Temperature:

0-50°C

Pressure Range:

0-20 bar

 

Parameters:

Oxygen [mg/l]

Sensor:

Derived parameters: oxygen saturation [%]

Clark electrode (amperometric)

Measuring range:

0-40mg

Measuring accuracy:

± 0.5% FS

Dissolution:

0.01 mg/l

Temperature:

0-50°C

Pressure Range:

0-10 bar

 

Alternative
Sensor:

Oxygen [mg/l]

Optical

Measuring range:

0-25mg/l (at 25°C, 1013hPa)

0-40mg/l (at 3°C, 1013hPa)

Measuring accuracy:

± 0.5% FS

Dissolution:

0.001mg/l

Temperature:

-5°C-50°C

Pressure Range:

0-12 bar

 

Sensor:

Turbidity [NTU]

Derived parameters: suspended solids [mg/l]

Optical

Measuring range:

0-1000 NTU and 0-5000 NTU

Measuring accuracy:

± 1% of the respective end of measuring range

Dissolution:

± 0.0125% of the respective end of measuring range

Temperature:

0-50°C

Pressure Range:

0-10 bar (with wiper)

0-20bar (without wiper)

 

Housing
Material:

1.4404 stainless steel (standard)

Dimensions:

48mm, length: 493mm

Protection class
Probe:

IP68

RS485 interface:

IP67 (immersion depth with dry cartridge up to 3m, depending on cable length)

 

Outputs:

RS485 (SHWP) standard, MODBUS, SDI-12, 4-20mA (Optional)

 

Communications
Interface:

RS485

 

Power supply:

4-15VDC or 8-15VDC with optical sensors

 

Operating temperature:

Parameter-dependent

 

Hanging depth:

up to 1000m

 

Cable:

Reinforced measuring cable with pressure compensation line

Gore-tex membrane, desiccant cartridge with SilicaGel for protection of the pressure compensation line